Anemometer(Tecpel AVM-715) Product ID: Tecpel AVM-715

LCD Digital Anemometer
AVM-715 Function:

ON/OFF, H/T, °F/°C, 10 Memories, Air Velocity Unit, AVG (Multipoint),
Air Flow Vlume Measurement, AVG/MAX/MIN Reading.



Features:

Instant Display of the air Velocity and temperature

Maximum , Minimum and average Value Calculation

Easy Air flow Volume calculation (CMF/CMM)

10 Memories Data Storage/Recalling/Clearing

Data HOLD/TEST function

Low Battery Indication

One Hour Auto Shut off function

Category: Anemometers
Certificate:CE
  • AVM-715 SPECIFICATION:
    • MEASUREMENT RANGE: -50°C ~ +70°C OR –58°F ~ +158°F
    • RESOLUTION : ± 0.1° FOR WHOLE RANGE
    • ACCURACY : ±0.5°C FOR 0°C ~ +50°C OR ±0.9°F FOR +32°F ~ +122°F, OTHERWISE ±1°C OR 1.8°
    • F
    • SENSING TIME: 1 SECOND
    • BATTERY: 3 X 1.5 VOLTS “AAA” SIZE OR EQUIVALENT
    • AIR VELOCITY MEASUREMENT :
    • MEASUREMENT RANGE RESOLUTION
    • 0.0 TO 45.0 m/s (METER PER SECOND) 0.1
    • 0.0 TO 140.0 km/hr (KILOMETERS PER HOUR) 0.1
    • 0.0 TO 100.0 mph (MILE PER HOUR) 0.1
    • 0 TO 8800 ft/min (FEET PER MINUTE) 10
    • 0.0 TO 88.0 knots (NAUTICAL MILES PER HOUR) 0.1
    • ACCURACY: ±2% FOR ALL RANGE
    • INSTALLATION:
    • UNPACK THE ANEMOMETER WITH CARE.
    • OPEN THE BATTERY DOOR. THEN INSTALL THREE 1.5 VOLTS “AAA” SIZE OR EQUIVALENT
    • BATTERY INTO THE BATTERY COMPARTMENT. AND CLOSE THE BATTERY DOOR.
    • REPLACE NEW BATTERY WHEN THE LOW BATTERY INDICATOR APPEARS ON THE SCREEN.
    • OPERATION:
    • ON/OFF
    • Press “ON/OFF” FOR ON/OFF SWITCHING. FULL DISPLAY SHOWS FOR A WHILE. THE UPPER
    • DISPLAY SHOWS THE VELOCITY (DEFAULT AS m/s) VELO SIGN APPEAR, WHILE THE LOWER
    • DISPLAY SHOWS THE TEMPERATURE (DEFAULT AS °F) MEASURED.
    • ONE HOUR AUTO SHUT OFF FUNCTION & CANCELLATION
    • THIS PRODUCT WILL TURN OFF AUTOMATICALLY AFTER ONE HOUR.
    • TO DISABLE THIS FUNCTION, PRESS “ON/OFF” AND “H/T” TOGETHER TO TURN ON THE UNIT
    • EACH TIME. THE PRODUCT WILL KEEP WORKING UNTIL BEING SHUT OFF BY USERS.
    • UNIT
    • THE UNIT IS DEFAULT IN ft/min AS MEASUREMENT UNIT FOR AIR VELOCITY. USER CAN PRESS
    • THIS BUTTON TO INTERCHANGE THE AIRFLOW VELOCITY UNIT AS THE FOLLOWING:
    • ft/min > mph > m/s > km/hr > knots.
    • HOLD/TEST
    • PRESS THE BUTTON TO FREEZE THE READING FOR RECORDING (BOTH VELOCITY &
    • TEMPERATURE), HOLD SIGN APPEARS.
    • PRESS AGAIN THE UNIT WILL RESTORE TO REAL TIME MODE.
    • °F/°C
    • PRESS TO SWITCH THE READING BETWEEN °C AND °F AS TEMPERATURE UNIT.
    • AVG/MAX/MIN READING FOR CONTINUE MOVING AVERAGE MEASUREMENT (SINGLE POINT)
    • a) PRESS ONCE THE UNIT WILL SHOW THE CONTINUE MOVING AVERAGE MEASUREMENT OF
    • VELOCITY (AVG), MAXIMUM MEASUREMENT (MAX) AND MINIMUM MEASUREMENT (MIN) AS
    • FOLLOWING:
    • AVG > MAX(BOTH TEMP & VELOCITY) > MIN(BOTH TEMP & VELOCITY) > REAL TIME READING
    • b) RELEVANT SIGN APPEARS ON THE DISPLAY AND READING UPDATE PER SECONDS.
    • c) TO CLEAR VALUE, PRESS AND HOLD THE KEY FOR 3 SECONDS, 2 SUCCESSIVE BEEPS
    • SOUND IS
    • GIVEN OUT.
    • AVG (MULTI POINT)
    • a) HOLD THE UNIT IN POSITION, PRESS THE “HOLD” BUTTON TO FREEZE THE READING (BOTH
    • VELOCITY AND TEMPERATURE).
    • b) THEN PRESS “MEM” BUTTON TO SAVE THE READING. (A SINGLE BEEP IS GIVEN OUT FOR
    • EVERY SUCCESSFUL SAVING). MEM 1 DISPLAYS REPRESENTING ONE READING IS SAVED.
    • c) REPEAT STEP (a) – (b) TO SAVE ANOTHER READING, MEM 12 DISPLAYS REPRESENTING 2
    • READINGS ARE SAVED.
    • d) CONTINUE TO DO SO UP TO 10 SETS OF READING CAN BE SAVED.
    • e) ONCE ALL MEASUREMENT HAVE BEEN SAVED, PRESS “AVG (MULTI POINT)” BUTTON, THE
    • UNIT WILL DISPLAY THE AVERAGE AIR VELOCITY FROM MEMORY AND AVG SIGN APPEARS.
    • MEM (MEMORY RECALLING)
    • a) Press THE “MEM” BUTTON, ALL SAVED READING (BOTH TEMPERATURE & VELOCITY) WILL
    • BE SHOWN, STARTING FROM THE 1ST READING (i.e. MEM 1) TO THE 10TH READING (i.e. MEM
    • 10).
    • b) PRESS AND HOLD THE “MEM” BUTTON FOR 3 SECONDS TO CLEAR ALL SAVED READINGS.
    • AIR FLOW VOLUME MEASUREMENT
    • (i) SINGLE POINT MEASUREMENT
    • a) DETERMINE THE FREE AREA OF THE AIR SOURCE.
    • b) PRESS “AIR FLOW” BUTTON FOR 3 SECONDS TO DEFINE FREE AREA, DISPLAYS SHOWS
    • AREA SIGN AND FIRST DIGIT OF 1.111 sq feet FLASHES.
    • c) PRESS “HOLD” BUTTON TO INCREASE NUMBER OF THE FLASHING DIGIT.
    • d) PRESS “AVG (MULTI POINT)” BUTTON TO MOVE TO THE NEXT DIGIT.
    • e) PRESS THE “AIR FLOW” BUTTON, THE DEFINED FREE AREA IS SAVED AND THE UNIT IS NOW
    • READY TO MEASURE AIR FLOW (CFM) AND FLOW SIGN APPEARS. VALUE SHOWS ON THE
    • UPPER DISPLAY.
    • (i) AVERAGE FOR MULTI POINT MEASUREMENT
    • PRESS AND HOLD THE “AVG (MULTI POINT)” BUTTON FOR 3 SECONDS TO CLEAR CURRENT
    • AVERAGE READINGS IN THE VELOCITY/FPM MODE.
    • a) REPEAT (a) TO (e) IN SECTION (G) TO DETERMINE ALL MULTI-POINT AVERAGE READING.
    • b) PRESS “AIR FLOW” BUTTON FOR 3 SECONDS AND UNIT ENTERS TO THE FREE AREA
    • SETTING.
    • c) REPEAT STEP (a) TO (e) IN SECTION (Ii) TO DEFAULT FREE AREA SETTING.
    • d) PRESS “AIR FLOW” BUTTON TO CONFIRM AND UNIT ENTERS TO THE CFM MODE. THE UNIT
    • DISPLAY AVERAGE AIR FLOW READING AND NUMBER OF POINTS MEASURED. VALUE SHOWS
    • ON THE UPPER DISPLAY.
    • PRECAUTIONS
    • REMOVE THE BATTERY FROM THE BATTERY COMPARTMENT IF THE ANEMOMETER WILL NOT
    • BE USED FOR LONG DURATION OF TIME OR DURING STORAGE.
    • THE THERMOMETER CANNOT BE USED OR PUT INSIDE ANY OVEN OR MICROWAVE.